Operation
8
Operation
8.1
Switching the Device On
and Off
Requirements
The device is connected to the power supply.
Procedure
CAUTION Pointed or sharp-edged objects may
t
damage the operating display!
Only touch the operating display with your fingertips.
t
To switch the device on: Press the [On/Off] button.
t
To switch the device off: Press and hold the [On/Off]
t
button.
8.2
Waiting for the Warm-up Time
In order to provide accurate weighing results, the device
must have reached the necessary operating tempera-
ture. The operating temperature is reached after the
warm-up time. The warm-up time starts when the device
is switched on.
Procedure
Switch on the device.
t
Wait until the operating temperature is reached
t
(warm-up time see Chapter "15.7 Warm-up Time",
page 38).
If a weighing process is carried out on a con-
formity-assessed device during the warm-up
time: The weight value is marked as invalid.
8.3
Leveling the Device with a
Level Indicator
Unevenness at the installation site of the device may re-
sult in incorrect weighing results. Leveling balances out
unevenness at the installation site.
26
Operating Instructions Gold&Carat II Line
Procedure
1
To level the device: Bring the air bubble of the level
t
indicator into the middle of the circular marking. To
do so, turn the leveling feet to the left or right.
Check that the device is standing on 4 leveling feet
t
and the support foot (1) is not in contact with the
setup surface.
8.4
Overview of Calibration and
Adjustment
During calibration, a calibration weight is used to
determine the deviation of the displayed value from the
actual value. The subsequent adjustment eliminates this
deviation.
We recommend regular calibration and adjustment:
— Daily, every time the device is switched on
— After every leveling
— After changing the ambient conditions (temperature,
humidity, or air pressure)
— After setting the device up at a new installation site
8.5
Calibrating and Adjusting
the Device Internally
(Only Model I-2x)
Requirements
— The weighing pan is unloaded.
— The operating display shows a stable weight value.
— The internal calibration and adjustment function is
set (see Chapter "7.2.1 Setting Internal Calibration
and Adjustment (Only Model I-2x)", page 16).
Unscrew the rear level-
t
ing feet, until they touch
the setup surface.
The support foot (1) is
y
not in contact with the
setup surface.