Descargar Imprimir esta página

Sartorius Gold & Carat II Line GL822I-2 Serie Instrucciones De Manejo página 17

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 129
7.3
Parameter List
7.3.1
"SETUP"/"BALANCE" Menu
Parameter
Setting values
AMBIENT
V.STABLE
STABLE*
UNSTABL.
V.UNSTBL.
APP.FILT.
FINAL.RD.*
FILLING
STAB.RNG
V. ACC.
ACC.*
FAST
V. FAST
* Factory setting
** Deviating factory setting: Only for models with x= SVN
*** Factory setting for gold and carat only
Explanation
Sets the ambient conditions to "very stable": Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
— Very stable table near the wall
— Closed and calm room
Sets the ambient conditions to "stable". Recommended for the following
work environment:
— Stable table
— Slight movement in the room
— Slight draft
Sets the ambient conditions to "unstable": Activates the delayed change in
weight values with a reduced output rate. Recommended for the following
work environment:
— Simple office desk
— Room with moving machinery or personnel
— Slight air movement
Sets the ambient conditions to "very unstable": Activates a significantly
delayed change in the weight values and long wait for stability with a further
reduction in the output rate.
Recommended for the following work environment:
— Noticeable and slow floor vibrations
— Noticeable building vibrations
— Weighed goods moved
— Very strong air movements
Activates a filter that enables a fast change in the display for very fast load
changes. Display changes with minimal load changes (in the digit range)
occur more slowly.
Activates a filter that enables a very fast change in the display with minimal
load changes, e.g. when filling containers.
Sets the stability to "very accurate".
Sets the stability to "accurate".
Sets the stability to "fast".
Sets the stability to "very fast".
System Settings
Operating Instructions Gold&Carat II Line
17

Publicidad

loading