Procedure
Press the [Zero] button.
t
Press the [Adjust] button.
t
If the calibration function with subsequent automatic
t
adjustment is selected:
The "CAL.RUN." display appears in the operating
y
display during the process.
The "CAL.END" display indicates the end of
y
calibration.
If the calibration function without subsequent
t
automatic adjustment is selected:
The "CAL.RUN." display appears in the operating
y
display during the process.
The calibration error is displayed.
y
Press the [Confirm] button.
t
The adjustment is performed.
y
The "CAL.END" display indicates the end of
y
calibration.
8.6
Externally Calibrating and
Adjusting the Device
Material
Calibration and adjustment weight
Requirements
— The weighing pan is unloaded.
— The operating display shows a stable weight value.
— The external calibration and adjustment function is
set (see Chapter 7.2.2, page 16).
— The preset weight value for the adjustment weight or
the user-defined weight value for the adjustment
weight is selected (see Chapter 7.2.2, page 16).
Procedure
Press the [Zero] button.
t
Press the [Adjust] button.
t
Place the indicated calibration and adjustment
t
weight on the balance.
The weight value is displayed.
y
If the calibration with subsequent automatic adjust-
t
ment function is selected and the calibration weight
placed on the device is within the specified limits:
The adjustment process starts.
y
If the "+" prefix was displayed: The applied weight
y
is too big.
If the "-" prefix was displayed: The applied weight
y
is too small.
The "CAL.END" message indicates the end of cali-
y
bration.
Remove the calibration weight.
t
If the calibration without subsequent automatic ad-
t
justment function is selected and the calibration
weight placed on the device is within the specified
limits:
The calibration process will start.
y
If the "+" prefix was displayed: The applied weight
y
is too big.
If the "-" prefix was displayed: The applied weight
y
is too small.
The calibration error is displayed.
y
Press the [Confirm] button.
t
The adjustment is performed.
y
The "CAL.END" message indicates the end of cali-
y
bration.
Remove the calibration weight.
t
8.7
Printing Results of the
Calibration and Adjustment
Process
The results of the calibration and adjustment process
can be printed in a GLP printout.
Procedure
Set the GLP printout in the menu (setting parame-
t
ters, see Chapter 7.3.7, page 22).
Calibrate the device.
t
Once the calibration is complete: The printing
y
process starts.
8.8
Weighing
Requirements
— The device has been leveled.
— The device is calibrated and adjusted.
NOTICE
Chemicals may damage the device or accessories!
Chemicals can attack the device or the connected
accessories internally and externally. This may damage
the device and accessories.
Use appropriate containers when weighing
t
chemicals.
Operating Instructions Gold&Carat II Line
Operation
27