15.10.2 Models GL3202 |GL822 | GL8201
Model
Scale interval (d)
Maximum capacity (Max)
Weighing system
Electromagnetic force compensation (EMFC)
Strain gauge (DMS)
Repeatability
At approx. the maximum load, typical value
Linearity deviation
Limits
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-2x models):
Temperature change
Time interval
Only for verified models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Typical measurement time
Recommended Calibration Weight
External calibrated test weight
Accuracy class in accordance with OIML R111-1
Weighing pan size
Net weight, approx.
Gross weight, approx.
* For the following models only: External, with verification
** For the following models only: External, without verification
Unit
Value
mg
10
g
3200
x
x
mg
10
± mg
20
± ppm/K
2*/3,5**
<100% of maximum capacity
K
2
h
6
II
BC-EF
mg
100
mg
500
s
≤1,0
g
2000
F1
mm
∅ 180
kg
4,4*/4,1** | 6,0
kg
5,2*/4,9** | 6,8
Operating Instructions Gold&Carat II Line
Technical Data
Value
Value
10
100
820
8200
x
x
x
x
10
100
20
300** / 100
2
7** / 2
2
2
6
6
II
II
BC-EF
BC_EF
100
100
500
5000
≤1,0
≤1,0
500
5000
F2
F2
∅ 180
∅ 180
4,4*/4,1** | 6,0
4,4*/3,9** | 6,0
5,2*/4,9** | 6,8
5,2*/4,7** | 6,8
41