Technical Data
15.10 Metrological Data
15.10.1 Models GL5202 | GL4202 | GL2202
Model
Scale interval (d)
Maximum capacity (Max)
Weighing system
Electromagnetic force compensation (EMFC)
Strain gauge (DMS)
Repeatability
At approx. the maximum load, typical value
Linearity deviation
Limits
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-2x models):
Temperature change
Time interval
Only for verified models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Typical measurement time
Recommended Calibration Weight
External calibrated test weight
Accuracy class in accordance with OIML R111-1
Weighing pan size
Net weight, approx.
Gross weight, approx.
* For the following models only: External, with verification
** For the following models only: External, without verification
40
Operating Instructions Gold&Carat II Line
Unit
Value
Value
mg
10
10
g
5200
4200
x
x
mg
10
10
± mg
20
20
± ppm/K
2
2
<100% of maximum capacity
K
2
2
h
6
6
II
II
BC-EF
BC-EF
mg
100
100
mg
500
500
s
≤1,0
≤1,0
g
5000
2000
F1
F1
mm
∅ 180
∅ 180
kg
4,4 | 6,0
4,4 | 6,0
kg
5,2 | 6,8
5,2 | 6,8
Value
100
2200
x
x
10
20
2*/3,5**
2
6
II
BC-EF
100
500
≤1,0
2000
F1
∅ 180
4,4*/4,1** | 6,0