Descargar Imprimir esta página

Sartorius Gold & Carat II Line GL822I-2 Serie Instrucciones De Manejo página 36

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 129
Technical Data
15 Technical Data
15.1
Ambient Conditions
Installation site: For indoor use only, max. height above sea level
Temperature
Environment (metrological data)*
Environment
Storage and shipping
Relative humidity**
At temperatures up to 31°C, non-condensing
Then linear decrease from max. 80% at 31°C to max. 50% at 40°C
No heat from heating systems or direct sunlight
No electromagnetic fields
* For conformity-assessed (verified) balances in accordance with EU requirements, refer to the information on the
balance.
** For conformity-assessed (verified) balances in accordance with EU requirements, the legal regulations apply.
15.2 Contamination Type, Overvoltage Category (Device)
Pollution level according to IEC 61010-1
Overvoltage category according to IEC 60664-1
15.3 Power Supply
15.3.1
Power Supply Device
Input voltage
Power consumption, max.
Only by Sartorius power supply unit YEPS01-15V0W
36
Operating Instructions Gold&Carat II Line
Unit
Value
m
3000
°C
+10 – +30
°C
+5 – +40
°C
-10 – +60
%
15 – 80
Unit
Value
2
I
Unit
Value
V
15 (±10%)
DC
W
4

Publicidad

loading