Durr Dental VistaScan Mini Instrucciones De Montaje página 85

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 151
Problem
Top or bottom bulge in the
X-ray image
X-ray image is mirror-in-
verted
Round shadow on the
X-ray image
Ghosting or double expo-
sure on X-ray image
X-ray image mirrored in
one corner
2141100009L30 1612V001
Probable cause
Image plate fed in off-centre and
at an angle
Image plate exposed on the
wrong side�
Plus ID image plate (with marker)
exposed on the wrong side
Image plate exposed twice
Image plate not sufficiently erased
Image plate bent during X-ray ex-
posure
Troubleshooting
Solution
Check the error code on the
i
touch screen�
Insert the image plate centrally
i
and straight�
Insert the image plate correctly
i
in the light protection cover�
Position the image plate cor-
i
rectly�
When taking an X-ray, make
i
sure that the active side faces
towards the X-ray tube�
Only expose the image plate
i
once�
Check the erasure unit is work-
i
ing correctly�
Inform a service technician if the
i
problem persists�
Do not bend the image plate�
i
EN
33

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Vistascan mini plus

Tabla de contenido