Telstar BiOptima Manual De Usuario página 11

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Features
External dimensions (LxWxH)
(Width including armrest)
Internal dimensions (LxWxH)
Front opening height
Weight
Downflow velocity
Inlet velocity /
Exhaust flow
Consumption*
Voltage
Frequency
Lighting
Noise
Vibration
HEPA/ULPA H 14 Filters
Fans
* Cabinet with factory settings and new filter.
DISSOCIATED DECLARED VALUES OF NOISE EMISSION
Standard(s) to which conformity is declared ISO 4871
L
(dB)
pA
K
(dB)
pA
Values measured according to acoustic test in standard EN 12469:2000 using standard
UM-BiOptima-EN-1907
Units.
mm
mm
mm
Kg
m/s
m/s
3
m
/h
W
V
Hz
Lux
dB (A)
mm RMS
V
downflow
BiOptima Cyto 4
≤60
2
ISO 11201
BiOptima Cyto 4
1338 x 832 x 2212
1190 x 548 x 666
285
0.35
0.6
510
283
≤ 60 (Attached table)
< 0.005
Efficiency 99.995%, EN 1822 by MPPS method
filtration system
(99.999% D.O.P. test @ 0.3µm particles)
High-efficiency fans.
Flows measured by thermoanemometric probes.
= 0.35m/s
V
= 0.6 m/s
inlet
(230 V, 50 Hz)
BiOptima Cyto 6
BiOptima Cyto 6
1948 x 832 x 2212
1800 x 548 x 666
200
380
0.35
0.6
773
420
230
50/60
≥ 1000
≤60
2
9 / 39

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Bioptima cyto

Tabla de contenido