Features
External dimensions (LxWxH)
(Width including armrest)
Internal dimensions (LxWxH)
Front opening height
Weight
Downflow velocity
Inlet velocity /
Exhaust flow
Consumption*
Voltage
Frequency
Lighting
Noise
Vibration
HEPA/ULPA H 14 Filters
Fans
* Cabinet with factory settings and new filter.
DISSOCIATED DECLARED VALUES OF NOISE EMISSION
Standard(s) to which conformity is declared ISO 4871
L
(dB)
pA
K
(dB)
pA
Values measured according to acoustic test in standard EN 12469:2000 using standard
UM-BiOptima-EN-1907
Units.
mm
mm
mm
Kg
m/s
m/s
3
m
/h
W
V
Hz
Lux
dB (A)
mm RMS
V
downflow
BiOptima Cyto 4
≤60
2
ISO 11201
BiOptima Cyto 4
1338 x 832 x 2212
1190 x 548 x 666
285
0.35
0.6
510
283
≤ 60 (Attached table)
< 0.005
Efficiency 99.995%, EN 1822 by MPPS method
filtration system
(99.999% D.O.P. test @ 0.3µm particles)
High-efficiency fans.
Flows measured by thermoanemometric probes.
= 0.35m/s
V
= 0.6 m/s
inlet
(230 V, 50 Hz)
BiOptima Cyto 6
BiOptima Cyto 6
1948 x 832 x 2212
1800 x 548 x 666
200
380
0.35
0.6
773
420
230
50/60
≥ 1000
≤60
2
9 / 39