No.
Type of
error
Open
frontal
Alarm
8
"
LIFTED
FRONTAL
"
(see Chapter
4.5 pto. D)
Light does
9
not turn on
UV
germicidal
10
lamp does
not turn on
Intermittent
11
alarm
activated
Display
12
does not
turn on
WARNING
All repair and maintenance work in the cabinet should be done by qualified technical staff with
knowledge about the maintenance procedures required by the cabinet.
UM-BiOptima-EN-1907
Causes of error
Lifted frontal inductive
sensor unadjusted.
Main control is damaged.
Light not working properly.
Closed frontal inductive
sensor unadjusted.
Glass not in closed position.
UV lamp does not work.
Ballast is damaged.
Inductive sensors are
unadjusted.
Main control is damaged.
Power cord cable
disconnected.
No power.
Main control is damaged.
Correction of error
Open the frontal and check that the open
frontal position inductive sensor is not
loose.
Notify the Technical Service to receive
instructions to adjust the inductive sensor.
Call a specialised technician to replace the
control.
Check that the connection in the control
card is correct (Annex A7).
Call a specialised technician to replace the
Led bar.
Open the frontal and check that the closed
frontal position inductive sensor is not
loose.
Notify the Technical Service to receive
instructions to adjust the inductive sensor.
Check the front slides properly.
Check the lamp holders are properly
connected.
Check that the connection in the control
card is correct (Annex A7).
Change the UV lamp.
Change the ballast.
Open the frontal and check that the
inductive sensors are not loose.
Notify the Technical Service to receive
instructions to adjust the inductive sensors.
Call a specialised technician to replace the
control.
Check that power cord cable is properly
connected in socket on the wall.
Check voltage on the socket of the wall.
Call a specialised technician to replace the
control.
39 / 39