No.
Type of
error
Laminar
down flow
not working
properly
4
Alarm "
LOW
LAM. FLOW
"
(see Chapter
4.5 point D)
Exhaust not
working
properly
5
Alarm "
LOW
INFLOW"
(see Chapter
4.5 point D)
Window
6
does not
move
Glass out of
position
Alarm
"
WORK
7
POSITION"
(see Chapter
4.5 pto. D)
UM-BiOptima-EN-1907
Causes of error
Probe not working properly.
Laminar down flow fan
stopped.
Exhaust completely covered
up.
Probe not working properly.
Exhaust fan stopped.
Front is blocked.
Motor of the window does
not work.
Inductive sensors
unadjusted.
Main control is damaged.
Glass position.
Working position inductive
sensors unadjusted.
Main control is damaged.
Correction of error
Check that the connection in the control card
is correct (Annex A7).
Check the fuse has not blown (Annex A7).
Check that the connections in the control
card are correct (Annex A7).
Notify the Technical Service to change the
fan.
Visually check the exhaust to remove any
type of obstruction.
Check that the connection in the control card
is correct (Annex A7).
Check the fuse has not blown (Annex A7).
Check that the connections in the control
card are correct (Annex A7).
Notify the Technical Service to change the
fan.
Visually check the front movement to remove
any type of obstruction.
Check the fuse has not blown (Annex A7).
Notify the Technical Service to change the
motor of the window.
Notify the Technical Service to adjust the
Inductive sensors position.
Call a specialised technician to replace the
control.
Move the glass into the working position.
Open the frontal and check that the working
position inductive sensor is not loose.
Notify the Technical Service to receive
instructions to adjust the inductive sensor.
Call a specialised technician to replace the
control.