VIII. Checking for Leaks
EN
After approximately one minute has elapsed the flow rate should be observed at the
drip chamber to ensure that the continuous flush device is operating properly. A visual
inspection for leaks should also be made since a small leak can misrepresent the actual
continuous flow through the catheter. A protocol should be established according to
the hospital standard of care for routinely checking the system for proper fluid source
pressure, flow rate and leaks.
IX.
Indications and Contraindications
A. Indications
● Direct arterial blood pressure monitoring
● Left atrial monitoring with an air-eliminating filter between solution source and
continuous flush device
● Pulmonary artery monitoring (PA distal)
● Venous pressure monitoring (RA proximal)
B. Contraindications
● Left atrial monitoring without an air-eliminating filter between solution source and
continuous flush device
● Intracranial pressure monitoring
● Compartmental pressure monitoring
● Intrauterine pressure monitoring
X.
SAFESET with In-Line Reservoir Indications and Contraindications
A. Indications
● To be used for blood withdrawals on patients that require periodic withdrawal
of blood samples. This system should only be used with a monitoring line that
includes a flush device.
B. Contraindications
● Not to be used on monitoring lines that do not contain a flush device.
THE TRANSPAC
®
4 KIT IS A SINGLE USE DEVICE. DO NOT REUSE. REUSE
NEGATIVELY IMPACTS PERFORMANCE / STERILITY POTENTIALLY RESULTING IN
PRODUCT FAILURE/ CONTAMINATION.