ACCURACY
In a study, 300 urine specimens collected from women for the purpose of
pregnancy testing, were evaluated with the Alere
TestPack hCG Urine and
TM
with the Clearview HCG Urine test.
Of the 300 urine specimens evaluated, 131 specimens tested positive by
both methods and 169 specimens tested negative by both methods. Thus an
agreement of 100% (300/300) was determined for these specimens. Therefore
the diagnostic sensitivity and diagnostic specificity of Alere
Urine and Clearview HCG were found to be comparable in this study. As with
all diagnostic tests, there is the potential for false positive or false negative
results to occur with this test.
The study results are summarised below:
TestPack
Clearview HCG Urine
hCG Urine
+
+
131
-
0
ADVICE LINE
For further information, please contact your distributor, or call Alere
Technical Specialists:
Africa, Russia & CIS:
Tel: +972 8 9429 683
Email: ARCISproductsupport@alere.com
Asia Pacific:
Tel: +61 7 3363 7711
Email: APproductsupport@alere.com
Europe & Middle East: Tel: +44 161 483 9032
Email: EMEproductsupport@alere.com
Latin America:
Tel: +57 01800 094 9393
Email: LAproductsupport@alere.com
Canada:
Tel: +1 800 818 8335
Email: CANproductsupport@alere.com
EN
TestPack hCG
TM
-
0
169