Descargar Imprimir esta página

Alarm Indicators - GCE Zen-O RS-00500 Manual Del Usuario

Ocultar thumbs Ver también para Zen-O RS-00500:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 172
4. To disable the Airplane Mode: Press and hold both Mute
for 4 seconds.
IMPORTANT: Activating or de-activating the Airplane mode does not affect the normal
operating performance of your device, continue to use your device as normal.

9 . ALARM INDICATORS

If the Zen-O™ Portable Oxygen Concentrator detects an alarm condition, it will indicate
the alarm visually and audibly within 10 seconds. There are four levels of alarms: critical
high priority, high priority, medium priority, and low priority.
Each is indicated differently by the backlit display; yellow, and red LEDs; and buzzer, as
indicated below. In each case, the alarm message and power status will override the cur-
rent display.
IMPORTANT: All alarm conditions and parameters are factory preset; conditions and
parameters cannot be changed or adjusted by the user.
IMPORTANT: The alarm system is tested during the startup sequence. You should see
all alarm lights briefly turn on and the audible alarm indicator chirp.
ALARM STATUS
Critical high
priority
High priority
Medium priority
Low priority
IMPORTANT: If two alarm conditions exist at the same time, the highest priority alarm is
indicated. If two or more alarm conditions of equal priority exist at the same time, the
most recent one will be displayed.
IMPORTANT: The most recent alarms indicated by the device are logged for reference
by service personnel. This log is maintained even if the device is powered down or if
power is lost for any other reason.
AUDIBLE TONE
Ten beeps per burst, a burst
repeats every 3 seconds.
Ten beeps per burst, a burst
repeats every 3 seconds.
Three beeps per burst,
burst repeats every
8 seconds
Three beeps per burst,
burst repeats every
10 minutes
25/338
and
VISUAL INDICATOR
Solid red LED and device is
automatically disabled
Flashing red LED
Flashing yellow LED
Solid yellow LED
Minus buttons
EN
MUTE TIME
20 minutes
20 minutes
8 hours
24 hours

Publicidad

Solución de problemas

loading

Este manual también es adecuado para: