Degree of protection
(water):
Technical Description
Degree of safety
(flammable anaesthetic
mixture):
Operating
temperature:
Storage temperature:
Alarm sound pressure
level:
Alarm system delays:
Oxygen concentrator
status indicator:
IP22 in carry bag (protection against small objects and tilted
dripping water)
IP20 out of carry bag (protection against small objects and no
protection against water entering the concentrator)
Not suitable for use in the presence of a flammable anaesthetic
mixture
Continuous operation at temperatures between 5°C (41°F) and
40°C (104°F).
Between -20°C (-4°F) and 60°C (140°F).
69 dB(A)
Less than 10 seconds after detection (low oxygen alarms if oxy-
gen is less than 82% volume fraction at specified environmental
conditions)
High priority alarm that indicates when oxygen concentration
drops below 82%
33/338