Instillation of fluids 7b
Fluids can be instilled through the working channel by inserting a syringe into the working
channel port at the top of the endoscope. When using a Luer Lock syringe, use the included
introducer. Insert the syringe completely into the working channel port or the introducer and
press the plunger to instill fluid. Make sure you do not apply suction during this process, as this
will direct the instilled fluids into the suction collection system. To ensure that all fluid has left
the channel, flush the channel with 2ml of air. It is recommended to remove introducer from
the working channel port when it is not in use.
Aspiration 7c
When a suction system is connected to the suction connector, suction can be applied by
pressing the suction button with the index finger. If the introducer and/or an endoscopic
accessory is placed inside the working channel note that the suction capability will be reduced.
For optimal suction capability it is recommended to remove the introducer or syringe entirely
during suction.
Insertion of endoscopic accessories 7d
Always make sure to select the correct size endoscopic accessory for the endoscope. Inspect
the endoscopic accessory before using it. If there is any irregularity in its operation or external
appearance, replace it. Insert the endoscopic accessory into the working channel port and
advance it carefully through the working channel until it can be seen on the aView. The
enclosed introducer can be used to facilitate the insertion of soft accessories such as microbi-
ology brushes.
Withdrawal of the endoscope 8
When withdrawing the endoscope, make sure that the control lever is in the neutral position.
Slowly withdraw the endoscope while watching the live image on the displaying unit.
4.3. After use
Visual check 9
1.
Are there any missing parts on the bending section, lens, or insertion cord?
If yes, then take corrective action to locate the missing part(s).
2.
Is there any evidence of damage on the bending section, lens, or insertion cord? If yes,
then examine the integrity of the product and conclude if there are any missing parts.
3.
Are there cuts, holes, sagging, swelling or other irregularities on the bending section,
lens, or insertion cord? If yes, then examine the product to conclude if there are any miss-
ing parts.
In case of corrective actions needed (step 1 to 3) act according to local hospital procedures.
The elements of the insertion cord are radio opaque.
Disconnect
Disconnect the endoscope from the displaying unit 10 .
Disposal
Dispose of the endoscope, which is a single-use device . The endoscope, is considered contam-
inated after use and must be disposed of in accordance with local guidelines for collection of
infected medical devices with electronic components.
5. Technical product specifications
5.1. Standards applied
The endoscope function conforms with:
– EN 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and
essential performance
– EN 60601-2-18 Medical electrical equipment - Part 2-18 Particular requirements for the basic
safety and essential performance of endoscopic equipment
10