To lower the foot end of the litter, press down on the pedal (D) or push down on the side of the uni-lower pedal (C) closest to
the foot end of the product.
P P o o s s i i t t i i o o n n i i n n g g t t h h e e p p r r o o d d u u c c t t i i n n T T r r e e n n d d e e l l e e n n b b u u r r g g
W W A A R R N N I I N N G G - - Always remove any devices that may be in the way before you raise or lower the litter.
C C A A U U T T I I O O N N - - Do not use the hydraulics on the base to raise the product with a patient lift under the product.
To position the product in the Trendelenburg position (head down), raise the litter to the highest height ( Raising the non-
electric litter (page 16) or Raising the electric litter option (page 17)).
N N o o t t e e - - Raise the litter to the highest height for a greater Trendelenburg angle.
To lower the head end of the product, push down on the head end or foot end release pedal (D) or push down on the side of
the uni-lower pedal (C) closest to the head end ( Prime base controls (page 16)).
To lower the product from Trendelenburg position, push down on the head end and foot end release pedals (B and D) at the
same time or push down on the center of the uni-lower pedal (C) until the litter is flat.
P P o o s s i i t t i i o o n n i i n n g g t t h h e e p p r r o o d d u u c c t t i i n n R R e e v v e e r r s s e e T T r r e e n n d d e e l l e e n n b b u u r r g g
W W A A R R N N I I N N G G - - Always remove any devices that may be in the way before you raise or lower the litter.
C C A A U U T T I I O O N N - - Do not use the hydraulics on the base to raise the product with a patient lift under the product.
To position the product in the Reverse Trendelenburg position (foot down), raise the litter to the highest height ( Raising the
non-electric litter (page 16) or Raising the electric litter option (page 17)).
N N o o t t e e - - Raise the litter to the highest height for a greater Trendelenburg angle.
To lower the foot end of the product, push down on the head end or foot end release pedal (B) or push down on the side of
the uni-lower pedal (C) closest to the foot end ( Prime base controls (page 16)).
To lower the product from Reverse Trendelenburg position, push down on the head end and foot end release pedals (B and
D) at the same time or push down on the center of the uni-lower pedal (C) until the litter is flat.
T T r r a a n n s s p p o o r r t t i i n n g g a a p p a a t t i i e e n n t t
W W A A R R N N I I N N G G
• Always position the patient in the center of the product.
• Always lock the siderails in the full up position with the sleep surface horizontal and in the lowest position when you
transport a patient.
• Do not transport the product laterally on inclines greater than 5.7 degrees (10% grade) to avoid product tipping.
To transport a patient:
1. Raise the siderails to the highest position. Pull on the siderail to make sure that it is latched.
2. Lower the litter to the lowest position.
3. Unplug the power cord from the wall outlet if your product is equipped with the electric lift or electric litter option.
4. Push down on the steer (green) side of the brake/steer pedal.
5. Use the push handles to push the product.