C C l l e e a a n n i i n n g g
C C l l e e a a n n i i n n g g t t h h e e p p r r o o d d u u c c t t
C C A A U U T T I I O O N N - - Do not use abrasive cleaners to clean the display enclosure for the scale system option. Do not allow cleaning
solutions or other fluids to pool on the display unit. Dry all surfaces after spills or cleaning.
These instructions provide recommended cleaning methods for the P P r r i i m m e e S S e e r r i i e e s s stretcher.
This product is power washable. The product may show some signs of oxidation or discoloration from continuous washing.
However, no degradation of the product's performance characteristics or functionality will occur due to power washing as
long as the proper procedures are followed.
Recommended cleaning method:
1. If the product is equipped with the electric litter or electric lift option, unplug the power cord from the wall outlet before
you transport or clean the product.
2. Remove the mattress from the product.
3. Follow the cleaning solution manufacturer's dilution recommendations.
4. Stryker recommends the standard hospital cart washer for power washing.
5. Dry the product. Do not place the mattress on the product until the product is dry.
6. Check functionality before you return the product to service.
Raise and lower the product
Lock and unlock the brake/steer pedal in both positions
Latch and unlatch the siderails
Raise and lower the Fowler backrest
Raise and lower the Gatch
Make sure all components have proper lubrication
Make sure all labels are intact
N N o o t t e e
• Direct skin contact with visibly soiled, permeable material may increase the risk of infection.
• Do not steam clean the product.
• Clean the base hood storage area.
• Clean the bottom of the brake pads to prevent wax or floor remnant buildup.
• Some cleaning agents are corrosive in nature and may cause damage to the product if you use them improperly. If
you do not rinse and dry the product well, a corrosive residue may be left on the surface of the product that could
cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your
warranty.
C C l l e e a a n n i i n n g g t t h h e e p p r r o o d d u u c c t t u u s s i i n n g g a a c c a a r r t t w w a a s s h h e e r r
Stryker recommends using a standard hospital surgical cart washer to wash the stretcher a maximum of once per year for
the life of the product.
To clean the product with a cart washer:
1. If the product is equipped with the electric litter or electric lift option, unplug the power cord from the wall outlet before
you transport or clean the product.
2. Remove the mattress prior to washing the product. Do not wash the mattress with the product.
3. Position the Fowler backrest at 45°.
4. Place the product in the full Reverse Trendelenburg position (foot end down).
EN
46