Descargar Imprimir esta página

Varex Imaging XRpad2 HWC-M Digital X-ray Manual De Usuario página 49

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 181
Accessories & Replacement Parts
Check to ensure the battery pack is clean, dry, and free of foreign contamination or debris.
6
If cleaning is necessary, see
instructions.
Ensure the battery charger is powered on.
7
Orient the battery pack to match the orientation of the battery charger, and slide the battery
8
pack firmly into the battery charger. Keep the battery pack in the battery charger until all the
four charge status LEDs maintain a solid green, indicating a full charge. To remove, lift and
slide the battery out of the battery charger.
WARNING
13.3
Installing the Battery Pack
When installing or changing the battery pack in the X-ray detector, remove the XRpad2
Protective Insert or the used battery pack from the X-ray detector battery slot (see
Removing the Battery Pack" on page
WARNING
To install the battery pack:
Fully support the X-ray detector to prevent it from dropping or slipping.
1
Check to ensure the battery compartment of the X-ray detector is clean, dry, and free of
2
foreign contamination or debris. If cleaning is necessary, see
tery Pack" on page 48
Check to ensure the battery pack is clean, dry, and free of foreign contamination or debris.
3
If cleaning is necessary, refer to
cleaning instructions.
The battery pack is keyed and can only be inserted into the X-ray detector in one orientation.
4
Align the orientation of the battery pack to match the orientation required on the X-ray
detector.
Insert the charged battery pack into the X-ray detector in the corresponding orientation, and
5
gently press on the end cap until the latches secure the battery pack inside the detector.
Push the Power button on the X-ray detector to power on.
6
Check the battery charge status on the X-ray detector. If the battery charge status shows
7
sufficient battery charge is present, the X-ray detector is ready for use. If the battery charge
status shows lower than desired battery charge level, replace the battery with a charged
battery.
46
"Section 13.7, Cleaning the Battery Pack" on page 48
Do not drop or hit the battery pack against hard objects as this may
cause a risk of damage to the battery pack, which may result in
exposure to the corrosive cell contents, fire, or explosion.
47).
There is risk of explosion, personal injury, or damage to the product
if the battery pack is replaced by a non-OEM approved component.
for cleaning instructions.
"Section 13.7, Cleaning the Battery Pack" on page 48
www.vareximaging.com
XRpad2 HWC-M Detectors User Manual
for cleaning
"Section 13.4,
"Section 13.7, Cleaning the Bat-
for

Publicidad

loading