Descargar Imprimir esta página

Varex Imaging XRpad2 HWC-M Digital X-ray Manual De Usuario página 54

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 181
XRpad2 HWC-M Detectors User Manual
Table 12
Overview of the XRpad Interface and Power Unit (Continued)
6
Trigger In Signal from Hand Switch (Prep/Expose)
7
Trigger In/Out I/F
8
XRpad Interface and Power I/O
9
TRIG IN Signal
10
TRIG OUT Signal
11
LAN Port to Imaging Workstation
12
USB Connector
13
Remote Detector Power
14
Functional Ground Connector
15
Potential Equalization Connector
16
Power AC connector (
17
AC Input LED (Green) (
18
AC Power Switch
Table 13
includes the specification for the XRpad Interface and Power Unit.
Table 13
Specification of the XRpad Interface and Power Unit
Electrical Specification
AC Input Voltage [16]
AC Frequency [16]
DC Output [8]
Trigger In Signal from Hand Switch [6]
Trigger Out Signal to Generator [5]
Trigger In Signal [7, 9]
Trigger Out Signal [7, 10]
DC Output 5PF [7]
Protection Class I:
Mechanical Specification
Size
Temperature Ranges
Operating
)
)
100 V - 240 V
50 Hz/60 Hz
13.5 V/5 A, 15 V/1 A
(voltage level is dependent of load)
5 V - 24 V/10 mA (SELV)
Same level as Trigger In Signal
5 V (SELV)
5 V
5 V/40 mA
260 mm (L) × 205 mm (W) × 50 mm (H)
10°C to 35°C
www.vareximaging.com