XRpad Interface and Power Unit
Table 13
Specification of the XRpad Interface and Power Unit (Continued)
Transportation/Storage
Relative Humidity
Operating
Transportation/Storage
Ingress Protection Rating
IP40 rated (protection against particles > 1 mm)
WARNING
WARNING
WARNING
15.1
Cleaning the XRpad Interface and Power Unit
If the XRpad Interface and Power Unit surface is dirty or dusty, clean it with a cleaning cloth
dampened with ethanol or a diluted neutral detergent. If you are using a disinfectant other than
those specified, we recommend you consult a specialist for the procedure for disinfection. Turn
off the XRpad Interface and Power Unit, and disconnect the AC power cable, detector power,
and detector communication tethered cables before cleaning.
WARNING
52
All external signals that are connected to the XRpad Interface and
Power Unit (especially PREP/EXPOSE and Trigger signals) should be
from a Separated or Safety Extra-Low Voltage (SELV) circuit. Ignoring
this warning may result in electric shock, which may result in severe
personal injury, death, or substantial product damage.
The detector must not be used in a situation where it controls the X-ray
dose delivered to the patient.
The trigger signals from the detector must not be the sole activation
means for X-ray delivery, and must be used in association with other
control inputs: for example, a hand switch, safety interlocks, and a
means of controlling the beam duration.
When the XRpad Interface and Power Unit is to be cleaned, be sure to
turn OFF the XRpad Interface and Power Unit and unplug all cables.
NEVER use thinner, benzine, acetone, or other flammable cleaning
agents.
Ignoring this warning may result in explosion, fire, or electric shock,
which may result in severe personal injury, death, or substantial
product damage.
www.vareximaging.com
XRpad2 HWC-M Detectors User Manual
–10°C to 70°C
10% to 90%
0% to 90%