XRpad2 HWC-M Detectors User Manual
17.5
Powering Off the X-ray Detector
The X-ray detector is powered off by holding down one of the following for more than
four seconds:
X-ray detector
XRpad Interface and Power Unit push button
Extended hand switch push button (wired mode)
17.6
Temperature Warning Limits
The XRpad2 HWC-M detector provides two temperature limit alerts. The Digital Radiography
Software (as configured by the system integrator) displays a temperature symbol in the detector
display informing the user of the temperature.
When the near thermal limit occurs (typically 40°C), immediately finish the exam and then
power down the detector to reduce the thermal load.
When the maximum thermal limit occurs (typically at 43°C), the detector will power off
automatically, as soon as the ongoing exam has been completed.
18.0
Workflow for Acquiring an Image
Figure 17
shows the procedure for acquiring an image after starting the Radiography
Imaging Software. Details of the Radiography Imaging Software and the X-ray generator are
described in their corresponding operation manuals.
(Figure 1
[5]) (wireless and wired mode).
www.vareximaging.com
Workflow for Acquiring an Image
(Figure 13
[1]) (wired mode).