Inspection and Maintenance
19.2
Monthly Inspection
Perform the following inspection at least once a month. If there is a problem, immediately
ask your establishment's safety department to contact your dealer, distributor, or device
manufacturer.
Inspection
Execute the Dark Noise
and X-ray Uniformity tests
(refer to the XRpad2 Digital
X-ray Detector Reference
Manual).
Make sure that the XRpad
Interface and Power Unit is
clean from dirt or dust.
Check the manufacturing
date of all battery packs to
ensure that their age is not
five years or older.
19.3
Yearly Inspection
Perform the following inspection at least once a year. If there is any problem, immediately ask your
establishment's safety department to contact your dealer, distributor, or device manufacturer.
Inspection
Execute the Dark Noise,
X-ray Uniformity, Bad
Pixel, and Resolution tests
(refer to the XRpad2 Digital
X-ray Detector Reference
Manual).
64
Result
Date
Date
/
/
Good/Bad Good/Bad Good/Bad If there are changes in
Good/Bad Good/Bad Good/Bad Use the instructions in
Good/Bad Good/Bad Good/Bad Replace the battery pack with
Result
Date
Date
/
/
Good/Bad Good/Bad Good/Bad If there are changes in
www.vareximaging.com
XRpad2 HWC-M Detectors User Manual
Remedy
Date
/
performance, acquire new
calibration files as described
in the Digital Radiography
Software Manual.
Contact your dealer,
distributor, or device
manufacturer if there is any
problem.
"Section 13.7, Cleaning the
Battery Pack" on page 48
cleaning.
a new one and recycle the old
one.
Remedy
Date
/
performance, acquire new
calibration files as described
in the Digital Radiography
Software Manual.
Contact your dealer,
distributor, or device
manufacturer if there is any
problem.
for