CAUTION
Independent user manipulation of system components
Falling due to breakage of load-bearing components or malfunction of the product.
►
Manipulations to the product other than the tasks described in these instructions for use are
not permitted.
►
The battery may only be handled by Ottobock authorised, qualified personnel (no replace
ment by the user).
►
The product and any damaged components may only be opened and repaired by authorised,
qualified Ottobock personnel.
CAUTION
Mechanical stress on the product
>
Falling due to unexpected product behaviour as the result of a malfunction.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
►
Do not subject the product to mechanical vibrations or impacts.
►
Check the product for visible damage before each use.
CAUTION
Use of the product when battery charge level is too low
Falling due to unexpected behaviour of the prosthesis because of changed damping behaviour.
►
Check the current charge level before use and charge the prosthesis if required.
►
Note that the operating time of the product may be reduced at low ambient temperatures or
due to ageing of the battery.
CAUTION
Risk of pinching in the joint flexion area
Injuries due to pinching of body parts.
►
Ensure that fingers/body parts or soft tissue of the residual limb are not in this area when
bending the joint.
CAUTION
Penetration of dirt and humidity into the product
>
Falling due to unexpected product behaviour as the result of a malfunction.
>
Falling due to breakage of load-bearing components.
►
Ensure that no solid particles or foreign objects can penetrate into the product.
►
The knee joint is weather-proof but not resistant to corrosion. Therefore, the knee joint
should not come into contact with salt water, chlorinated water or other solutions (such as
soap or shower gel, and body and/or wound fluids). Do not use the knee joint under extreme
conditions like diving or jumping into water. The knee joint is not designed for prolonged
underwater use or prolonged submersion.
►
After contact with water, remove the Protective Cover (if installed) and hold the prosthesis
with the sole of the foot facing up until the water has drained from the knee joint and tube
adapter. Dry the knee joint and components with a lint-free cloth and allow the components
to fully air dry.
53