LED on
LED on bat
power
tery charger
supply
Beep signal
4 x short at intervals
of approx. 20 sec.
(continuously)
16.2.3 Error messages while establishing a connection with the cockpit app
Error message
Component was con
nected to another
device. Establish con
nection?
Mode change failed
Error
Battery is fully charged (or con
nection with product is interrup
ted).
Error
Charging the battery outside the
allowable temperature range
Cause
The component was
connected to another
device
An attempt was made to
switch to a different
MyMode while the com
ponent was in motion
(e. g. while walking)
A current connection to
the component was
interrupted
Resolution
Take note of the confirmation sig
nal for differentiation.
When the battery charger is con
nected or disconnected, a self-
test is conducted and confirmed
by a beep and vibration signal.
The battery is fully charged if this
signal is heard.
If no signal is emitted, the con
nection to the product is interrup
ted.
If the connection to the product is
interrupted, an authorised
Ottobock Service Centre must
inspect the product, battery char
ger and power supply.
Resolution
Check whether the specified ambi
ent conditions for charging the bat
tery are met (see Page 78).
Correction
To disconnect the original connection, tap
the "OK" button.
If the original connection is not to be dis
connected, tap the "Cancel" button.
For safety reasons, switching MyModes is
only permitted when components are at
rest, e. g. while standing or sitting.
Check the following points:
•
Distance from the component to the
device
•
Charge level of the component's bat
tery
•
Bluetooth of the component switched
on? ()
•
Hold the component with the sole of the
foot facing up to make the component
"visible" for 2 minutes.
•
If multiple components were stored,
was the correct component selected?
83