Withdraw damaged covers from service and replace.
Check zip fasteners for function and integrity.
Platform
Regularly inspect the interior and exterior of the platform, ensuring that the poppers and air cells are intact.
Ensure the platform has not drifted out of position and the air hoses are not blocked, kinked or twisted.
Fitting and Replacing Damaged Air Cells
In the event of an air cell puncture or leak, either or both air cells can be replaced. Please contact Frontier
Therapeutics Customer Services for replacement parts Tel: +44 (0) 330 460 6030.
1.
Switch off the control unit and disconnect from mains power.
2.
Disconnect the CPR air hoses from the foot end of the platform using the red quick connectors.
3.
Unzip and pull back the cover revealing the platform.
4.
Determine the faulty air cell and remove by unfastening the popper fastenings.
5.
* Disconnect the air hoses at the head end of the platform using the yellow and blue quick connectors.
6.
Unthread the CPR air hoses and red quick connectors from their securing loops at the foot end of the
platform.
7.
* Unthread the air hoses with the yellow and blue quick connectors at the head end of the platform from
their respective securing loops.
8.
Inspect the platform for other signs of damage, including the potential source of the damage.
9.
Fit the new air cell ensuring it is in-line with the correct section, and in the correct orientation with the
red CPR air hose quick connectors at the foot end of the platform. Secure in place with the popper
fastenings.
10.
Thread the CPR air hose and red quick connector through its securing loop at the foot end of the
platform.
11.
* Thread the air hose connections through their respective securing loops at the head end of the platform
and reconnect using the yellow and blue quick connectors.
12.
Re-fit the cover and close the zip ensuring the hose with the red quick connectors that attach to the CPR
air hoses are fed through the opening at the foot end of the cover.
13.
Re-attach the CPR air hoses at the foot end of the platform using the red quick connectors, ensuring
they are not twisted or kinked.
* Only applicable for Toto® Cradle.
Nature and Frequency of Preventative Maintenance
Inspection and Care of the Toto® Touch Control Unit
The Toto® Touch control unit is designed to be reliable and long lasting, with few parts requiring
maintenance.
A routine service is required when the service light illuminates. Please refer to the Toto® Service,
Maintenance and Repair Manual for further information. Supporting videos are also available.
Clean air filters at least every 12 months by removing and running under clean water and allowing to dry.
Inspections should be undertaken to note any visible signs of damage.
In the event of product failure, please contact Frontier Therapeutics Limited for advice and
recommendations on service and repair.
Although there are no requirements for Portable Appliance Testing, guideline recommendations indicate
that extension leads and portable electrical equipment should be re-tested every six months. Please refer to
local guidance.
An electrical safety test in accordance with BS EN 62353:2014 Medical Electrical Equipment - Recurrent Test
and Test After Repair of Medical Electrical Equipment, should be carried out after maintenance, inspection,
servicing and repair, and before the control unit is sent out to, or returned to a user.
Cleaning and Disinfection
All components of the system can be cleaned with detergent and water to remove any visible
contamination and chemically disinfected with chlorine. The following represents guidance on the correct
cleaning and decontamination process but does not replace local policy and guidelines.
Toto® Lateral Turning System Platform Cover
Keeping a cleaning record is recommended as the Toto® system is suitable for re-use.
Light and Heavy Soiling
Decontaminate with a 0.1% (1,000ppm) - 1% (10,000ppm) Chlorine solution.
Rinse with clean water and wipe down with a single-use, non-abrasive cloth and thoroughly dry.
Do not apply a 1% chlorine-based solution for more than two minutes, doing so may cause long term
damage.
Prolonged use of alcohol-based cleaning agents may reduce the life of the product.
If used, thoroughly rinse with clean water and allow to dry before use.
Machine Washing Instructions
Cleaning and disinfection may be achieved by laundering at temperatures not exceeding 65ºC
for ten minutes or 73ºC for three minutes.
12