Instructions for Use
Before use, please read the following instructions. Should you have any queries, please contact Frontier
Therapeutics Limited on +44 (0) 330 460 6030 or visit our website www.frontier-group.co.uk.
Platform Installation and Operating Guide
To ensure the system is correctly installed and functions smoothly, follow this step-by-step guide.
1.
Remove all items from packaging and inspect for damage.
2.
Remove the mattress and place the rolled platform at the foot end of the bed. Roll out until it is fully
extended and positioned centrally to the bed frame (A).
3.
Inspect the platform for damage and ensure the air cells are secured in position via the popper fastenings
by unzipping the cover and visually inspecting (B).
4.
* Ensure that the air hoses are connected by visually inspecting their yellow and blue quick connectors
(C).
5.
If required, re-fit the cover and close the zip ensuring the CPR air hoses are fed through the opening at
the foot end of the cover. Please Note: The platform cover should be free to move, allowing each side
of the platform unrestricted room to inflate.
6.
Connect the air inflation and air cell hoses using the red quick connectors, ensuring that they are not
twisted or kinked (D).
7.
Check that the CPR emergency deflation valve is securely closed and there are no air leaks (E).
8.
Replace the mattress on top of the platform. DO NOT strap or fasten the mattress to the bed frame.
9.
Using the bed hooks, hang the control unit on the bed foot board, ensuring it is securely in place (F).
10.
Reattach the CPR air hoses at the foot end of the platform to the control unit using the black quick
connectors, ensuring they are not twisted or kinked (G).
11.
Connect the control unit to mains power via the lead supplied (H).
* Only applicable for Toto® Cradle.
A. Remove the mattress and
place the rolled platform at the
foot end of the bed. Roll out until
it is fully extended and positioned
centrally to the bed frame.
C. Ensure that the air hoses are
connected by visually inspecting
their yellow and blue quick
connectors.
E. Check that the CPR
emergency deflation valve is
securely closed and there are no
air leaks.
G. Reattach the CPR air hoses at
the foot end of the platform to
the control unit using the black
quick connectors, ensuring they
are not twisted or kinked.
Patient Positioning
Position supine patients along the centre line of the mattress by aligning the patient's nose with the centre of
the headboard. Ensure the head and shoulders are supported preventing neck rolling and pivoting at the hips.
Toto® is indicated for side lying if patients can self-support, and they are in the middle of the bed.
When using bedside, consider the use of padding, bumpers, or net infills.
Centre Line
8
B. Inspect the platform for
damage and ensure the air
cells are secured in position
via the popper fastenings
by unzipping the cover and
visually inspecting.
D. Connect the air inflation
and air cell hoses using
the red quick connectors,
ensuring that they are not
twisted or kinked.
F. Using the bed hooks, hang
the control unit on the bed
foot board, ensuring it is
securely in place.
H. Connect the control unit to
the mains power via the lead
supplied.