1.3
Safety instructions
Operating manual uni-cart, vexio-cart, pro-cart, duo-cart, compact-cart, classic-cart and endo-cart 03/2014
TD_JEST_15_GA_001
Rev.: 001
General
• Only those mobile equipment carriers may be operated whose main voltage equipment
has been tested and approved by appropriate, qualified personnel!
• Personnel (hospital and service personnel) working directly or indirectly with a mobile
equipment carrier must be instructed!
• Setting adjustments may only be carried out by qualified personnel.
• Repairs may only be carried out by ITD GmbH.
Operation
• With every change of location it must be ensured that no one is injured or objects damaged!
Connections
• In the USA and Canada, use a hospital-grade connection cable when connecting isolating
transformers in the 115 V mode and in Japan use the optional Japanese connection cable.
• When connecting to the multiple-socket strip, secure the plug using an optional available plug
pull-out lock.
• Only equipment satisfying the requirements of IEC 60601-1 may be connected to the
sockets / connecting cables!
• Additional medical equipment with connecting bolts for equipotential bonding are to be
connected with the green-yellow cable to the optional equipotential bonding connecting bolts!
• Caution: The overall rating given on the nameplate must not be exceeded.
Please observe that no further multiple sockets must be connected to an
existing multiple socket.
Load capacity
• The total weight of the equipment and the accessories on the mobile equipment carrier may
not exceed the permitted payload weight (see load capacity sticker on the base frame).
• The surface load imprinted on the system components may not be exceeded!
• The load shown on the fittings (e.g. infusion stand, jointed arms) may not be exceeded!
Infection protection
• Hygiene regulations are to be observed when cleaning!
• Give only cleaned and disinfected equipment and fittings to a service technician for
maintenance and repair work!
Environmental protection
• Dispose of all cleaning and disinfection agent residue in a manner not harmful to the
environment!
Erstellt am/durch: 07.04.2015/ITD