9
Maintenance
Operating manual uni-cart, vexio-cart, pro-cart, duo-cart, compact-cart, classic-cart and endo-cart 03/2014
TD_JEST_15_GA_001
Rev.: 001
The uni-cart, vexio-cart, pro-cart, duo-cart, compact-cart, classic-cart or endo-cart has been
developed and constructed for many years of trouble-free use.
Check the functional capability of the following parts every 12 months in order to guarantee
safety.
Monitor shelf:
• Swivels and tilts easily without too much play.
Shelves:
• Check whether the mounting screws have been tightened and whether the shelf is stable
and flat.
Castors:
• Check that the castors run freely and the brakes operate correctly.
• Check that the 4 bolts holding the castors to the bottom side of the base and the castors
themselves are seated firmly in their holders.
• The running surfaces of castors must be free from contamination, so that they can function
correctly.
Socket strips:
• Check the main cable for damage and firm seating.
Auxiliary sockets:
• Check the cable for damage and firm seating.
Variable height support arms (flexion-port):
• The height adjustment functions freely, the raising force is adjusted to the weight of the
device.
Support arms:
• Rotating and tilting functions smoothly, without excess play.
Isolating transformers:
• Safety-relevant checks of isolating transformers.
Serial number:
• Compare the serial number of the cart with the data of the equipment log book.
Fuses:
• Check whether the correct fuses have been installed.
If you encounter any problems during these checks you should contact your supplier immediately.
Erstellt am/durch: 07.04.2015/ITD
Freigabe am/durch: 07.04.2015/ETI
Maintenance