Electrical safety
3.9
EMC
3.10 Excluded from the final electrical inspection of system components
and accessories
40
TD_JEST_15_GA_001
Rev.: 001
If you have any questions, you should contact your local dealer or technical services.
Please note:
Make sure that this is also the case for adaptation of equipment in the power supply circuit
(e. g. multiple socket strip).
Electromagnetic compatibility of the medical electrical equipment positioned on the cart has
to be checked by the overall system configurer. If you use a different equipment combination,
you should check electromagnetic compatibility between individual items of equipment before
using in a medical setting.
The carrier system cannot be used in an Magnetic Resonance environment due to the
presence of ferro-magnetic materials.
ITD GmbH excludes any liability in this respect.
Using the carrier system with an isolating transformer within an environment subject to
explosive hazard is not permissible.
ITD GmbH exclude the following system components and accessories from the final electrical
inspection:
• Multiple sockets that are not wired in the assembly
• ME cables and appliance cables included
• POAG plates and cables included