2.2 Bench Alignment Using an Alignment Tool such as L.A.S.A.R. / PRO.S.A. Assembly

INFORMATION
To align the prosthesis please proceed in two steps:
First make the bench alignment using an alignment tool such as 743L200 L.A.S.A.R. Assembly / 743A200
PRO.S.A. Assembly.
In a second step, the static alignment is optimized with the 743L100 L.A.S.A.R. Posture (see Section 2.5 "Static
Alignment Optimization").
A
+5 mm
A correct bench alignment (e.g. using the 743L200 L.A.S.A.R. Assembly / 743A200 PRO.S.A. Assembly.) ensures
that the user can benefit from all the advantages of the C-Leg. The optimal residual limb position must be anticipat-
ed when positioning the socket connector. Plumb lines in the frontal and sagittal planes (drawn from the hip joint's
centre of rotation and marked during plaster cast taking and trial fitting of the test socket) will facilitate correct
positioning of the lamination anchor or socket adapter.
• Position the middle of the foot (MF) approx. 30 mm anterior to the alignment reference line (A). This applies to all
feet that are recommended for use with the C-Leg, independently of the specifications in the instruction manuals
of those feet!
Add 5 mm to the required heel height. Set correct outward rotation of the foot (Fig. 1).
• Clamp the knee joint with the mounted tube adapter into the alignment tool. Place the alignment reference
point (=knee axis) approx. 0-5 mm anterior to the alignment reference line. Consider the knee-floor distance
and outward rotation of the knee (approx. 5° are provided for by the adapter insert in the L.A.S.A.R. Assem-
bly / PRO.S.A. Assembly.). Recommended positioning of the alignment reference point: 20 mm above the medial
tibial plateau (Fig. 2).
• Connect the foot to the modular knee joint using a tube adapter. To do so, tilt the joint in the correct position and
set the required tube length (Fig. 2).
• Mark the centre of the socket proximally (M) and distally on the lateral side. Draw a line through both marks from
the socket brim to the distal end of the socket (Fig. 3).
Now position the socket such that the alignment reference line (A) passes through the proximal centre mark (M).
Set the socket flexion to somewhere between 3° and 5°; however, the individual situation (e.g. hip joint contrac-
38 Ottobock
MF
30 mm
5 mm
1
A
M
3
2
M
A
+ 5 mm
2
C-Leg Prosthetic System
3°- 5°
MF
4