Material
The connector tool consists of:
• Body and knob made of molded CELCON®
• Collet made of PEEK (polyetheretherketone)
• Two spring contacts made of MP35N®
Storage
Store in a cool, dark, dry place. Recommended storage temperatures are -30°C to 60°C.
INSTRUCTIONS FOR USE
WARNING: Use caution handling the lead terminal when the connector tool is not present on the lead. Do
not directly contact the lead terminal with any surgical instruments or electrical connections such as PSA
(alligator) clips, ECG connections, forceps, hemostats, and clamps. This could damage the lead terminal,
possibly compromising the sealing integrity and result in loss of therapy or inappropriate therapy, such as
an electrical short circuit within the header.
Carefully read all instructions prior to use.
1.
Remove any stylet from the lead.
2. Slide the connector tool onto the proximal end of the lead.
3. Pinch the terminal boot levers and continue sliding the connector tool until the terminal boot molding
aligns with the indicator arrows to ensure the lead is fully inserted. (Figure 2)
4.
Release the terminal boot levers to secure the connector tool to the proximal end of the lead.
5. Select a stylet and insert it through the stylet funnel into the lead. (Figure 3)
1 CELCON is a registered trademark of CNA Holdings LLC.
2 M90 is a trademark of Celanese Corporation.
3 MP35N is a registered trademark of SPS Technologies, Inc.
C
D
E
Figure 1.
(acetal copolymer) M90™
1
3
A
B
D
Figure 2.
2
A Stylet funnel
B Fixation knob
C Cathode (-) spring contact
D Anode (+) spring contact
E Indicator arrows
F Terminal boot levers
G Rotation mark
2
A Terminal boot levers
B Indicator arrows
C Terminal boot molding
D Terminal molding
C