Descargar Imprimir esta página

Replacing Hook Tip - Ottobock 8E600 Instrucciones De Uso

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 55
Pronation, supination procedure
CAUTION
Risk of pinching between the hook tips
Injury due to pinching of body parts.
Ensure that no body parts are between the hook tips when using or rotating the product.
INFORMATION
The following steps apply only to the 9S501 AxonRotation adapter (mechanical rotation). When
using the 9S503 AxonRotation, pronation and supination are controlled by the associated
muscle signals.
9 Cleaning
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der­
maClean).
1) Clean the product with a damp cloth and mild soap (e.g. Ottobock 453H10=1 Derma Clean)
when needed.
Ensure that no liquid penetrates into the system component(s).
2) Dry the product with a lint-free cloth and allow it to air dry fully.
10 Maintenance

10.1 Replacing hook tip

Required Tools
711M93 cap nut mounting wrench
9S305 hook tip (set)
Flexible mode
1) Push the unlock lever of the Axon-Bus gripping component (to the
stop) until it engages.
→ The wrist unit can be moved without engaging at the ratchet
positions.
2) When the unlock lever is pushed again, the wrist unit engages at
the next available position.
Rigid mode
1) Push the unlock lever of the Axon-Bus gripping component lightly
(not to the stop) and move the wrist unit to the desired position.
2) When the unlock lever is released, the wrist unit engages at the
next available position.
The wrist unit supports pronation and supination.
► Grasp the gearbox and turn the wrist unit to the desired position.
Ottobock | 23