The electrodes are to be placed on intact skin only and with as much electrode-skin contact
as possible. In the case of strong interference from electronic devices, the position of the
electrodes should be checked and changed if necessary. If the interference cannot be elim­
inated or if you do not achieve the expected results by adjustment or selection of the appro­
priate control programme, please contact the Ottobock branch responsible for your country.
Set the electrode gain as low as possible in order to reduce interference from powerful elec­
tromagnetic radiation (e.g. visible or or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong electromagnetic interference (e.g. high-voltage lines, transmitters, trans­
former stations, computer tomographs, magnetic resonance tomographs, etc.).
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
Only clean the product and the contacts with a damp cloth and mild soap (e.g. 453H10=1
Ottobock DermaClean).
INFORMATION
If you do not achieve the results expected by adjusting and selecting the appropriate control
mode, please contact Ottobock Myo-Service.

5 Scope of Delivery and Accessories

Scope of Delivery
1 pc. 8E600=* AxonHook
1 pc. 647G1001 instructions for use (qualified personnel)
Accessories
The following components are not included in the scope of delivery and must be ordered separ­
ately:
1 pc. 757L500 AxonCharge Integral
"560X500=* AxonSoft" adjustment software
1 pc. 757B500/757B501 AxonEnergy Integral
1 pc. 13E500 AxonMaster (firmware 2.0.70 or higher)
1 pc. 9S503 AxonRotation (active rotation unit)
or
1 pc. 9S501 AxonRotation adapter (passive rotation unit)

6 Charging the prosthesis battery

The following points must be observed when charging the battery:
The capacity of a fully charged battery is sufficient for one full day.
We recommend charging the product overnight when used by the patient on a daily basis.
When used daily, the complete charging unit (power supply – battery charger) may remain
plugged into the wall socket.
The battery should be charged for at least 3 hours prior to initial use.

6.1 Connecting the power supply and battery charger

1) Slide the country-specific plug adapter onto the power supply until it locks into place.