Charging recept­
acle
-
-
Error while charging the product
Battery charger

12.2.3 Status signals

Battery charger is connected
Battery charger
26 | Ottobock
Beeper
-
3 x short
Pulsating
Explanation/correction
Axon-Bus prosthetic system is not charging
Disconnect and then reconnect the charging plug.
Clean charger contacts.
Charger is overheated or overcooled
Disconnect charging plug and allow charger to cool
down/warm up.
Battery charger or power supply unit is defective
Consult prosthetist
Explanation
Power supply and battery charger operational
Battery is charging
Battery is charging, battery 25% charged
Battery is charging, battery 50% charged
Explanation/correction
Axon-Bus prosthetic system outside operat­
ing temperature range
Allow the Axon-Bus prosthetic system to
warm up or cool down (Technical data)
Axon-Bus prosthetic system switches itself
off
Charge battery
Opening the Axon-Bus gripping component
in an emergency
System is switched off, system can be used
without restriction after being switched on
again