Vascular Applications Only - Medtronic SYNCHROMED II 8637 Manual De Implantación

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Filename Date Time
UC200xxxxxx EN
4.6 x 6.0 inches (140 mm x 197 mm)
Template
Figure 11. Open clamp and inject into catheter access port.
8. During injection, check the needle puncture site for swelling.
9. When the drug has been injected, close the clamp, and carefully remove the syringe.
10. Proceed to "Flushing the catheter access port" on page 32.

Vascular applications only

1. Using sterile procedures, assemble the needle, extension tubing, filter, and syringe
containing the prescribed fluid.
2. Purge the air from the fluid pathway.
3. Close the clamp and gently insert the needle into the catheter access port septum until
the needle touches the needle stop.
4. Open the clamp and inject directly at an infusion rate not greater than 5 mL per minute
(Figure 12).
w
Warning: During vascular applications, do not aspirate blood through the catheter
access port or catheter. Blood sampling or aspiration through the catheter access
port is contraindicated in vascular applications. Residual blood from aspiration or
blood sampling can occlude the catheter or pump and inhibit drug delivery,
resulting in a loss of or change in therapy, which may lead to a return of underlying
symptoms, drug withdrawal symptoms, or a clinically significant or fatal drug
underdose, and require surgical revision or replacement.
M221311A028 Rev A
Template version: 03-03-2010
Syringe with prescribe d fluid
Open clamp
Needle
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
8637 2009-10
English 31
2009-10

Publicidad

Tabla de contenido
loading

Tabla de contenido