Current Drain At Beginning Of Life; Correspondence Between Magnet Rate And Battery Impedance - sorin REPLY DR Manual De Implante

Marcapasos cardiaco bicameral con adaptación de frecuencia
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

10.1. CURRENT DRAIN AT BEGINNING OF LIFE

Shipment conditions (1)
Shipment conditions (1)
Cenelec conditions (2)
Cenelec conditions (2)
-1
(1) 60 min
, 3.5 V, 0.35 ms, 500 Ohms, Holter ON, sensors ON.
-1
(2) 70 min
, 2.5 V, 0.5 ms, 500 Ohms, Holter ON, sensors ON.
(3) Twenty minutes after the detection of implantation, the device
switches automatically from DDD to safeR mode, assuming 1%
ventricular pacing and 100% atrial pacing.
10.2. CORRESPONDENCE BETWEEN MAGNET
RATE AND BATTERY IMPEDANCE
Magnet rate (min
Magnet period (ms)
Battery resistance (kilohms)
Magnet rate (min
Magnet period (ms)
Battery resistance (kilohms)
Magnet rate (min
Magnet period (ms)
Battery resistance (kilohms)
safeR mode (3)
DDD mode
safeR mode (3)
DDD mode
)
96
-1
625
<3
)
89
-1
672
7.5
)
78
-1
773
11
Inhibited: 6.5 µA
100% pacing: 10.4 µA
Inhibited: 6.5 µA
100% pacing: 14.2 µA
Inhibited: 6.9 µA
100% pacing: 9.6 µA
Inhibited: 6.9 µA
100% pacing: 12.3 µA
95
94
93
633
641
648
3
4
5
88
86
83
680
695
719
8
8.5
9
75
73
70
797
820
859
12
13
>14
91
90
656
664
6
7
82
80
734
750
9.5
10
– 35
ENGLISH