Special features
Smoothing
Acceleration (%)
AV delay shortening (ms)
Mode switching
Anti-PMT protection
Atrial or ventricular Autosensing
Ventricular Auto-threshold
Minimum ventricular amplitude (V)
Post V Atrial Blanking (ms)
(1)
After ventricular pacing, the minimal blanking in atrium is equal to the
programmed value. After ventricular sensing, the minimal blanking in
atrium is equal to the programmed value minus 55 ms.
AA prevention functions
Pause suppression
Acceleration on PAC
Overdrive
Maximum Overdrive rate (min-1)
Values
OFF-Very slow-Slow-Medium-Fast
0-5-15-25-35-45
0-15-30-45-65-80-95-110
ON-OFF
Termin-Reprog
Auto-Monitor
Auto-Monitor-OFF
1.5-2-2.5-3.0-3.5 (± 10 %)
150-165-180-195-210-225-240-255
(1)
(± 10 ms)
Values
A-V-A+V-OFF
ON-OFF
ON-OFF
100-110-130-155-185 (± 15 min-1)
– 39
ENGLISH