Pacing/Sensing
Atrial or ventricular amplitude (V)
Atrial or ventricular pulse width
(ms)
(1)
Atrial sensitivity (mV)
Ventricular sensitivity (mV)
Atrial or ventricular sensing polarity
(5)
Atrial or ventricular pacing polarity
(5)
(1)
The nominal value may be greater than the underlined value (return to
programmed value).
(2)
Values are measured using a positive and negative triangular signal of 2/13
ms.
(3)
For sensitivity settings below 2.5 mV, the pacemaker may detect noise lower
than the level specified in clause 27.5.1 of standard EN 45502-2-1.
(4)
For sensitivity settings below 0.3 mV, the pacemaker may detect noise
lower than the level specified in clause 27.4 of standard EN 45502-2-1
(5)
As soon as the detection of implantation has been confirmed, the lead
configuration is automatically programmed to unipolar pacing and
bipolar sensing (if a bipolar lead is used).
38 –
ENGLISH
(1)
(2)(3)(4)
(2)(3)
Values
1.5-2.0-2.5-3.0-3.5-4.0-5.0-7.5
(± 20 %)
0.10-0.25-0.35-0.50-0.60-0.75-
0.85-1 (± 35 µs)
0.1-0.2-0.3-0.4 (± 0.025 mV ±
40 %)
0.6-0.8-1.0-1.2-1.5-1.8-2.0-2.2-2.5-
2.7-3.0-3.5-4.0-4.5-5.0-6.0
(± 20 %)
1.0-1.2-1.5-1.8-2.0-2.2-2.5-2.7-3.0-
3.5-4.0-4.5-5.0-6.0-8.0-10.0-15.0
(± 20 %)
Unipolar-Bipolar
Unipolar-Bipolar