Errors And How To Clear Them; Cleaning; Observe The Following - Kulzer cara Print LEDcure Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 69

8.8 Errors and how to clear them

DISCONNECT FROM MAINS
Risk of life-threatening consequences warning
Unplug from the power supply before working on the unit.
ELECTRIC SHOCK
High voltages are life-threatening when touched. There are no components inside the unit that are
serviceable or repairable by the user which require opening up the unit. Therefore, users are prohibited
from opening up the unit.
GB
Error
-18-
The unit does not
switch on
Polymerization does not
start or is cancelled with
error message
Item will be very hot
9

Cleaning

ELECTRIC SHOCK
The LED modules work with low voltage but high current. This presents a relevant electrical hazard.
If moisture enters by way of cleaning and disinfecting agents, there is a hazard of short circuit and
electric shock.

Observe the following:

Switch off the unit and unplug the power supply before cleaning.
Disassembling the housing is not part of the operator's maintenance work.
Do not use spray bottles – only use slightly moistened wipes.
Indicator lights
none
none
none
none
'Door open' ERR
3 x 500 ms
'LED defect' ERR
3 x 500 ms
'LED temperature exceeded'
3 x 500 ms
ERR
'Temp. sensor defect' ERR
3 x 500 ms
'Fan signal fail' ERR
3 x 500 ms
none
none
No error message
none
Signal tone
Cause
Unit disconnected
Fuse defect
Mains cable defect
Incorrect mains voltage
Door switch triggered
One or several LED module
defects
LED temperatur > 65°C
Chamber temperature
sensor defect
Fan defect
Air intake or output
obstructed
Ambient temperature too
high
Check polymerization
parameters for customer
programs
How to clear
Establish mains connection
Check fuse in the mains
connection plug on the rear
side of the unit and replace,
if necessary. If the error
reoccurs, remove the unit
from operation and notify
Service.
Replace mains cable
Check permitted input voltage
(see section 12 'Technical
details')
Close door correctly.
Re-execute polymerization,
if necessary.
Contact Service.
Let the unit cool down;
check air supply and fan
function (bottom and back);
check ambient conditions
(if necessary, air-condition
the room, ventilate)
Contact Service
Contact Service
Check the unit environment
or base, keep air intake and
output free of obstructions
Air-condition the room /
ventilate
Set of parameters according
to the manufacturer specifi-
cations. Reduce LED power
and temperature and extend
time, if necessary.

Publicidad

Tabla de contenido
loading