Keep the following in mind when measuring resistance.
NOTE
– The test leads can add 0.1 Ω to 0.2 Ω of error to resistance measurements. To
– Be aware that the resistance function can produce enough voltage to
Keysight U1241C/U1242C User's Guide
test the leads, touch the probe tips together and read the resistance of the
leads. To remove lead resistance from the measurement, hold the test lead
tips together and press
subtracted from all future display readings.
forward-bias silicon diode or transistor junctions.
Vsense
. Now the resistance at the probe tips will be