Table 2-8
Range
100 Ω
1000 Ω
10 kΩ
100 kΩ
1000 kΩ
10 MΩ
100 MΩ
You can set the beeper to sound and the red LED indicator to light up as a
NOTE
continuity indication whether the circuit-under-test is less than (short) or more
than or equal to (open) the threshold resistance. Refer to "Changing the
continuity type" on page 131.
– Normal open: The circuit is normally open, the beeper will sound and the LED
– Normal closed: The circuit is normally closed, the beeper will sound and the
The audible continuity will be locked in the minimum range of 1000 Ω for
resistance measurement.
– The continuity function detects intermittent shorts and opens lasting as
NOTE
– You can enable or disable the audible and visual alerts via the multimeter's
Keysight U1241C/U1242C User's Guide
Threshold resistance values
Resolution
Accuracy
0.01 Ω
0.2% +5
0.1 Ω
0.2% +2
0.001 kΩ
0.2% +2
0.01 kΩ
0.2% +2
0.1 kΩ
0.2% +2
0.001 MΩ
0.8% +2
1.5% +3 (< 50 MΩ)
0.01 MΩ
3.0% +3 (> 50 MΩ)
indicator will light up when a short is detected.
LED indicator will light up when an open is detected.
briefly as 1 ms. A brief short or open causes the multimeter to emit a short
beep and its red LED indicator to light up.
Setup. Refer to "Changing the continuity alert type" on page 132.
Making Measurements
Continuity threshold Overload protection
28 ± 10 Ω
28 ± 10 Ω
0.151 ± 0.05 kΩ
1000 Vrms
1.38 ± 0.5 kΩ
< 0.3 A short circuit current
13.8 ± 4.3 kΩ
0.12 ± 0.04 MΩ
0.12 ± 0.04 MΩ
2
67