In case the measure can not be transmitted, QuickRead returns the string I
Example :
Request of the measure status on QuickRead with selected metric system, high
resolution, UpperTolerance = +00.1000 and measure = +00.1341
Request from Host
Response from QuickRead
Example :
Request of the measure status on QuickRead with selected metric system, high
resolution, LowerTolerance = -00.1000 and measure = -00.0282
Request from Host
Response from QuickRead
M+S?
Request from Host
Response from QuickRead
Request of the Measure Status and Value.
If the measure is available, QuickRead returns the string xxxxxxxx consisting in the
formatted measure as described in chapter 9 and appends a character describing the status:
<
measure lower than the lower tolerance limit.
>
measure greater than the upper tolerance limit.
@
measure within the tolerance limit.
X
measure tolerance limits not available.
In case of overrange of the measure, QuickRead returns the string O
In case the measure is not zeroed, QuickRead returns the string N
In case the measure can not be transmitted, QuickRead returns the string I
Example :
Request of the measure status and value on QuickRead with selected metric
system, high resolution, UpperTolerance = +00.1000 and measure = +00.1341
Request from Host
Response from QuickRead
Example :
Request of the measure status and value on QuickRead with selected metric
system, high resolution, LowerTolerance = -00.1000 and measure = -00.0282
Request from Host
Response from QuickRead
Z
Request from Host
Response from QuickRead
Zeroing.
The QuickRead checks and performs the zeroing and returns .
In case of error QuickRead returns the string ERR3
The QuickRead display highlights the zeroing results.
S?
>
S?
@
M+S?
xxxxxxxx
.
.
M+S?
+00.1341>
M+S?
-00.0282@
Z
.
.
.