The pushbutton box features four Cannon 9-pin RS232 connectors as follows:
- Male connector for connection of up to four Quick Read
- Male connector for connection of the box to a PC
- Female connector for connection of the power supply cable
- Female connector for connection of the footswitch
3.11.3
Performing remote zeroing with the pushbutton box
1.
Connect the Quick Read to the box F (code 6139013100) through the serial cable E (code
6739696157) if only one unit is used, through the serial cable D (code 6739696396) if
more units (max. 4) are used
2.
Connect the power supply cable G (code 6739696301) from Quick Read to the box (from
the connector 2 or 3 of the last unit powered)
3.
Switch the Quick Reads on
4.
Enter the programming mode
5.
Program SYST. ID of each Quick Read as a unique number
6.
Program for each Quick Read COM.MODE = ascii or QRbin. (all of the units must have
the same programming)
7.
Push the Zeroing button of the box
ZEROED is displayed first on each unit, then the measuring value 0.000
"hold" measuring mode
3.11.4
1.
Connect the Quick Read to the box F (code 6139013100) through the serial cable E
(6739696157) if only one unit is used, through the serial cable D (code 6739696396) if
more units (max. 4) are used
2.
Connect the power supply cable G code 6739696301 from Quick Read to the box (from the
connector 2 or 3 of the last unit powered)
3.
Switch the Quick Reads on
4.
Enter the programming mode
5.
Program for each Quick Read M. TYPE = hold
6.
Program SYST. ID of each Quick Read as a unique number
7.
Program for each Quick Read COM.MODE = ascii or QRbin. (all of the units must have the
same programming)
8.
Push the hold button of the box
the displayed measuring value is "frozen"
9.
to return to measuring mode push the hold button again
3.11.5
Performing dynamic measuring mode with the pushbutton box
1.
Connect the Quick Read to the box F (code 6139013100) through the serial cable E
(6739696157) if only one unit is used, through the serial cable D (code 6739696396) if
more units (max. 4) are used.
Chapter 3. Installation