HOLD is displayed first, then the "frozen" measuring value blinking.
To return to measuring mode press ESC again
MEASURE is displayed first, then the current measuring value.
The "hold" measuring mode can be performed also through the Quick Read external pushbutton
box (optional) code 6139013100 (see paragraph 3.11 Accessories).
DYNAMIC MEASURING MODE ["Max", "min", "M – m", "(M + m)/2", "(M – m)/2"]
This mode allows to carry out different measurement elaborations according to the programmed
type.
To enable it proceed as follows:
Program the required measurement type [M. TYPE = "Max", "min", "M – m", "(M + m)/2",
1.
"(M – m)/2"], then exit the programming. In this condition Quick Read still elaborates a
static measurement
2.
To enable the programmed dynamic measurement type press DYN for two seconds
 DYNAMIC is displayed blinking (until the acquisition start command is given) and the
analog visualization disappears
3.
Press + to command the acquisition start
 START is displayed first, then the measuring value
4.
Press again + to stop the acquisition
 STOP is displayed first, then the "frozen" measuring value blinking
5.
To reset the measurement and re-start the acquisition press +
6.
To exit the dynamic mode and return to the static mode press ESC.
NOTE
It is possible to exit the dynamic mode only after having stopped the acquisition (STOP must
have been displayed).
NOTE
Alarm 83 is displayed if + is pressed to start the dynamic acquisition before having enabled it
by pressing DYN.
The dynamic measuring mode can be performed also through the Quick Read external
pushbutton box (optional) code 6139013100 (see paragraph 3.11 Accessories).