Descargar Imprimir esta página

Masimo RD SET Serie Manual página 7

Ocultar thumbs Ver también para RD SET Serie:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 29
Inf sensor for INFANTS (3–10 kg)
2. Refer to Fig. 2a. Direct the sensor cable so that it runs along the top of the foot. Position the detector on the fleshy pad of
the great toe. Alternatively, the toe next to the great toe, or the thumb can be used (not shown).
3. Refer to Fig. 2b. Wrap the adhesive wrap around the toe so the emiter is positioned on the nailbed of the great toe.
Complete coverage of the detector window is needed to ensure accurate data.
4. Refer to Fig. 2c. Ensure that the emitter window ( ) aligns on the top of the toe directly opposite the detector. Verify
correct positioning and reposition if necessary.
Neo sensor for NEONATES (< 3 kg) and NeoPt/NeoPt-500 sensor for PRETERMS (< 1 kg)
2. Refer to Fig. 3a. For fragile skin, the stickiness of the medical grade adhesive can be diminished or eliminated by daubing
the adhesive areas with a cotton ball or gauze.
3. Refer to Fig. 3b. Direct the sensor cable toward the ankle (or wrist) so that it runs along the top of the foot (or hand). Apply
the sensor around the foot (or hand) so that emitter and detector are aligned. Complete coverage of the detector window
is needed to ensure accurate data.
4. Refer to Fig. 3c. Wrap the adhesive/foam wrap around the foot (or hand) and ensure that the emitter window ( ) aligns
directly opposite of the detector. Be careful to maintain proper alignment of the detector and emitter windows while
attaching adhesive/foam wrap to secure the sensor.
5. Refer to Fig. 3d. Verify correct positioning and reposition if necessary.
Neo sensor for ADULTS (> 40 kg) Inf Sensor for INFANTS (10–20 kg)
2. Refer to Fig. 4a. Direct the sensor cable so that it runs along the top of the hand. Position the detector on the fleshy part
of the finger. Alternatively, the sensor may also be applied to the toe (not shown).
3. Refer to Fig. 4b. Wrap the adhesive wrap around the finger so the emitter window ( ) aligns on the top of the finger
directly opposite the detector. Complete coverage of the detector window is needed to ensure accurate data.
4. Refer to Fig. 4c. Check the sensor to verify correct positioning and reposition if necessary.
C) Attaching the Sensor to the Patient Cable
1. Refer to Fig. 5a. Orient the sensor's connector tab so that the side with the "shiny" contacts is facing up. Orient the patient
cable with the color bar and finger grips facing up.
2. Refer to Fig. 5b. Insert the sensor tab into the patient cable until there is a tactile or audible click of connection. Gently tug
on the connectors to ensure a positive contact. Tape may be used to secure the cable to the patient for ease of movement.
D) Reattachment
• The sensor may be reapplied to the same patient if the emitter and detector windows are clear and the adhesive still
adheres to the skin.
• If the adhesive no longer adheres to the skin, use a new sensor.
NOTE: When changing application sites, or reattaching sensor, first disconnect the sensor from the patient cable.
E) Disconnecting the Sensor from the Patient Cable
1. Refer to Fig. 6. Pull firmly on the sensor connector to remove it from the patient cable.
NOTE: To avoid damage, pull on the sensor connector, not the cable.
7
10497A-eIFU-1220

Publicidad

loading