Initial Device Set-up
Mains connection
EN
Grounding contact
Only for U.S. operators
Potential equalization
Fig. 4-1
Device front
ON/OFF switch
(1)
Monitor with touch screen
(2)
Footswitch connection
(3)
Gas heater connection
(4)
Cassette for Smoke evacuation
(5)
Connection for insufflation tube
(6)
set
Eject button
(7)
18
Ensure the connection data and power supply comply with DIN VDE or national
requirements. The mains connection cable may be plugged only into a properly
installed, grounded safety wall socket (shockproof socket) (see DIN VDE 0100-
710). Read the device label located in rear of device (type plate) to determine the
operating voltage of the device.
CAUTION!
Check to make sure the available mains voltage matches the data listed on the
label attached to the back of the device. Incorrect voltage can cause errors and
malfunctions and may destroy the device.
The power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 or IEC
60320. Grounding will only be reliable if the equipment is connected to a corre-
sponding hospital grade socket.
The equipotential bonding is used as a protective measure against the failure of
the protective conductor according to requirements of IEC 60601-1 in the respec-
tively valid version. The installation must be according to the relevant local safety
regulations.
4.3
Front of Device
(1)
(2)
Familiarize yourself with the control and function elements at the front of the de-
vice.
(3)
(4)
(5)
(6)
(7)