Descargar Imprimir esta página

Aesculap Flow 50 Instrucciones De Manejo página 148

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Electromagnetic Compatibility
EN
66
11.5
Recommended Separation Distances between Portable and Mo-
bile RF Communications Equipment and the Flow 50
The Flow 50 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user/operator of the insufflator
can help prevent electromagnetic interference by maintaining a minimum dis-
tance between portable and mobile RF communications equipment (transmit-
ters) and the insufflator as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
Separation distance according to frequency of transmitter
output power of
(m)
transmitter (W)
150 kHz to
80 MHz outside
ISM bands
d = 1.2P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recom-
mended separation distance d in metres (m) can be determined using the equa-
tion applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects and peo-
ple.
WARNING!
Portable HF communication devices can have an effect on the performance char-
acteristics of the device. Therefore such devices must be kept a minimum dis-
tance of 30 cm (independent of any calculation) from the device Flow50 and its
accessories and cables
80 MHz to
800 MHz to
800 MHz
2.7 GHz
d = 1.2P
d = 2.3P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23

Publicidad