Annual Inspection
EN
62
10.7
Gas Flow Rate Test
• Connect a tube set.
• Select a nominal gas flow rate of 15 l/min.
• Start insufflation:
Press the START button.
• Press the gas volume button (0.0 l must be displayed).
Now start measuring for one minute.
• Stop the insufflation after one minute: Press the STOP button.
The gas consumption should be at least 11 l.
Each successfully conducted test must be documented with the test log.