Descargar Imprimir esta página

Aesculap Flow 50 Instrucciones De Manejo página 99

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Initial Device Set-up
4.2
Device Setup and Connection
Place the device on a flat surface free of vibration located in a dry environment.
Setting up
The ambient temperature and humidity must meet the requirements mentioned
in Chapter 13 Technical Data.
WARNING!
EN
ME System (Medical Electrical System)
Use only parts and/or devices from ME systems (see Chapter 15 Glossary) in pa-
tient environments in compliance with the standard IEC60601-1 in the respec-
tive currently valid version.
CAUTION!
ME Device in Rack
If it should be necessary to operate the device close to stacked with other devic-
es, the ME device or ME system (see chapter 15 Glossary) should be monitored to
ensure it works properly as configured.
CAUTION!
ME systems
The medical electrical (ME) device is suitable for integration in ME equipment
systems (see Chapter 15 Glossary). Operation of the ME device in vicinity of non-
ME devices may result in voiding the intended use of the ME device.
CAUTION!
Equipment should be positioned such that power cord can be easily disconnect-
ed.
CAUTION!
Ventilation of the device
Avoid device overheating. Ensure free air circulation especially to the bottom
and rear of the device (rear panel distance of at least 10 cm).
CAUTION!
Position the device in such a way that it is easy to operate and switch off.
CAUTION!
Place the device outside the sterile field.
CAUTION!
User position
To avoid a malfunction, the user must be positioned correctly
• within a display viewing angle of ±50° to operate the device
• up to 2 m from the device front for monitoring the actual values
17

Publicidad

loading

Este manual también es adecuado para:

Pg150