9.4
Replacing the Fuse
WARNING!
Replace the fuse
Replace the fuse only with a fuse provided by the manufacturer (see chapter 14
Accessories).
The fuse may be defective and is in need of replacement if:
• displays and LEDs do not light up,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the power supply input
and to a safety socket,
• the fuse of the building's electrical supply is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does not have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply. To do so, pull the mains plug from the
wall socket.
3. Remove power connection cable from mains socket.
4. The fuse holder is located right next to the mains socket.
5. Remove fuse holder as depicted in Fig. 9-1 Opening the fuse holder.
6. A Undo the latch of the fuse holder with your fingers.
7. B Remove the fuse holder.
8. C Check fuse.
9. Insert a new fuse. Use only the specified type of fuse (see chapter 13 Techni-
cal Data).
10. Insert the fuse holder until it can be heard snapping into place.
11. Reconnect device with the mains power. To do so, plug the mains plug into
the wall socket.